B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
B. Braun recommends applying updates:
SpaceCom: Version U62 or later (United States), L82 or later (outside the United States)
Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States)
Data module compactplus: Version A12 or later
Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Configurations
Workarounds
As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:
Ensure the devices are not accessible directly from the Internet.
Use a firewall and isolate the medical devices from the business network.
Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Exploits
Credits
Julian Suleder, Nils Emmerich, and Birk Kauer of ERNW Research GmbH; Dr. Oliver Matula of ERNW Enno Rey Netzwerke GmbH, reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices).
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.