NVD Vulnerability Details :
CVE-2019-18256
Analyzed
More InfoOfficial Page
Source-ics-cert@hq.dhs.gov
View Known Exploited Vulnerability (KEV) details
Published At-29 Jun, 2020 | 14:15
Updated At-29 Oct, 2021 | 19:58

BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.

CISA Catalog
Date AddedDue DateVulnerability NameRequired Action
N/A
Metrics
TypeVersionBase scoreBase severityVector
Primary3.14.6MEDIUM
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Primary2.02.1LOW
AV:L/AC:L/Au:N/C:P/I:N/A:N
CPE Matches

biotronik
biotronik
>>cardiomessenger_ii-s_gsm_firmware>>2.20
cpe:2.3:o:biotronik:cardiomessenger_ii-s_gsm_firmware:2.20:*:*:*:*:*:*:*
biotronik
biotronik
>>cardiomessenger_ii-s_gsm>>-
cpe:2.3:h:biotronik:cardiomessenger_ii-s_gsm:-:*:*:*:*:*:*:*
biotronik
biotronik
>>cardiomessenger_ii-s_t-line_firmware>>2.20
cpe:2.3:o:biotronik:cardiomessenger_ii-s_t-line_firmware:2.20:*:*:*:*:*:*:*
biotronik
biotronik
>>cardiomessenger_ii-s_t-line>>-
cpe:2.3:h:biotronik:cardiomessenger_ii-s_t-line:-:*:*:*:*:*:*:*
Weaknesses
CWE IDTypeSource
CWE-522Primarynvd@nist.gov
CWE-257Secondaryics-cert@hq.dhs.gov
Evaluator Description

Evaluator Impact

Evaluator Solution

Vendor Statements

References
HyperlinkSourceResource
https://www.us-cert.gov/ics/advisories/icsma-20-170-05ics-cert@hq.dhs.gov
Third Party Advisory
US Government Resource
Change History
0Changes found

Details not found