Logo
-

Byte Open Security

(ByteOS Network)

Log In

Sign Up

ByteOS

Security
Vulnerability Details
Registries
Custom Views
Weaknesses
Attack Patterns
Filters & Tools
Vulnerability Details :

CVE-2022-32537

Summary
Assigner-Medtronic
Assigner Org ID-3c337f33-b2bd-4529-9792-f48cdb2838b4
Published At-17 Nov, 2022 | 20:47
Updated At-29 Apr, 2025 | 14:44
Rejected At-
Credits

Medtronic MiniMed 600 Series Pump System Communication Issue

A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance

Vendors
-
Not available
Products
-
Metrics (CVSS)
VersionBase scoreBase severityVector
Weaknesses
Attack Patterns
Solution/Workaround
References
HyperlinkResource Type
EPSS History
Score
Latest Score
-
N/A
No data available for selected date range
Percentile
Latest Percentile
-
N/A
No data available for selected date range
Stakeholder-Specific Vulnerability Categorization (SSVC)
▼Common Vulnerabilities and Exposures (CVE)
cve.org
Assigner:Medtronic
Assigner Org ID:3c337f33-b2bd-4529-9792-f48cdb2838b4
Published At:17 Nov, 2022 | 20:47
Updated At:29 Apr, 2025 | 14:44
Rejected At:
▼CVE Numbering Authority (CNA)
Medtronic MiniMed 600 Series Pump System Communication Issue

A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance

Affected Products
Vendor
Medtronic
Product
Minimed 600 Series Insulin Pump
Default Status
unaffected
Versions
Affected
  • 620G, 630G, 640G, 670G
Problem Types
TypeCWE IDDescription
CWECWE-693CWE-693
Type: CWE
CWE ID: CWE-693
Description: CWE-693
Metrics
VersionBase scoreBase severityVector
3.14.8MEDIUM
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
Version: 3.1
Base score: 4.8
Base severity: MEDIUM
Vector:
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
Metrics Other Info
Impacts
CAPEC IDDescription
Solutions

Configurations

Workarounds

Exploits

Credits

Timeline
EventDate
Replaced By

Rejected Reason

References
HyperlinkResource
https://www.cisa.gov/uscert/ics/advisories/icsma-22-263-01
N/A
https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-600-series-communication-issue.html
N/A
Hyperlink: https://www.cisa.gov/uscert/ics/advisories/icsma-22-263-01
Resource: N/A
Hyperlink: https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-600-series-communication-issue.html
Resource: N/A
▼Authorized Data Publishers (ADP)
1. CVE Program Container
Affected Products
Metrics
VersionBase scoreBase severityVector
Metrics Other Info
Impacts
CAPEC IDDescription
Solutions

Configurations

Workarounds

Exploits

Credits

Timeline
EventDate
Replaced By

Rejected Reason

References
HyperlinkResource
https://www.cisa.gov/uscert/ics/advisories/icsma-22-263-01
x_transferred
https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-600-series-communication-issue.html
x_transferred
Hyperlink: https://www.cisa.gov/uscert/ics/advisories/icsma-22-263-01
Resource:
x_transferred
Hyperlink: https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-600-series-communication-issue.html
Resource:
x_transferred
2. CISA ADP Vulnrichment
Affected Products
Metrics
VersionBase scoreBase severityVector
Metrics Other Info
Impacts
CAPEC IDDescription
Solutions

Configurations

Workarounds

Exploits

Credits

Timeline
EventDate
Replaced By

Rejected Reason

References
HyperlinkResource
Information is not available yet
▼National Vulnerability Database (NVD)
nvd.nist.gov
Source:security@medtronic.com
Published At:12 Dec, 2022 | 13:15
Updated At:07 Nov, 2023 | 03:47

A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance

CISA Catalog
Date AddedDue DateVulnerability NameRequired Action
N/A
Date Added: N/A
Due Date: N/A
Vulnerability Name: N/A
Required Action: N/A
Metrics
TypeVersionBase scoreBase severityVector
Primary3.14.8MEDIUM
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
Secondary3.14.8MEDIUM
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
Type: Primary
Version: 3.1
Base score: 4.8
Base severity: MEDIUM
Vector:
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
Type: Secondary
Version: 3.1
Base score: 4.8
Base severity: MEDIUM
Vector:
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
CPE Matches

medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7730_firmware>>-
cpe:2.3:o:medtronic:guardian_link_2_transmitter_mmt-7730_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7730>>-
cpe:2.3:h:medtronic:guardian_link_2_transmitter_mmt-7730:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7731_firmware>>-
cpe:2.3:o:medtronic:guardian_link_2_transmitter_mmt-7731_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7731>>-
cpe:2.3:h:medtronic:guardian_link_2_transmitter_mmt-7731:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7738_firmware>>-
cpe:2.3:o:medtronic:guardian_link_2_transmitter_mmt-7738_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7738>>-
cpe:2.3:h:medtronic:guardian_link_2_transmitter_mmt-7738:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7775_firmware>>-
cpe:2.3:o:medtronic:guardian_link_2_transmitter_mmt-7775_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_2_transmitter_mmt-7775>>-
cpe:2.3:h:medtronic:guardian_link_2_transmitter_mmt-7775:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_3_transmitter_mmt-7810_firmware>>-
cpe:2.3:o:medtronic:guardian_link_3_transmitter_mmt-7810_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_3_transmitter_mmt-7810>>-
cpe:2.3:h:medtronic:guardian_link_3_transmitter_mmt-7810:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_3_transmitter_mmt-7811_firmware>>-
cpe:2.3:o:medtronic:guardian_link_3_transmitter_mmt-7811_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>guardian_link_3_transmitter_mmt-7811>>-
cpe:2.3:h:medtronic:guardian_link_3_transmitter_mmt-7811:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_620g_mmt-1750_firmware>>-
cpe:2.3:o:medtronic:minimed_620g_mmt-1750_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_620g_mmt-1750>>-
cpe:2.3:h:medtronic:minimed_620g_mmt-1750:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_630g_mmt-1715_firmware>>-
cpe:2.3:o:medtronic:minimed_630g_mmt-1715_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_630g_mmt-1715>>-
cpe:2.3:h:medtronic:minimed_630g_mmt-1715:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_630g_mmt-1754_firmware>>-
cpe:2.3:o:medtronic:minimed_630g_mmt-1754_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_630g_mmt-1754>>-
cpe:2.3:h:medtronic:minimed_630g_mmt-1754:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_630g_mmt-1755_firmware>>-
cpe:2.3:o:medtronic:minimed_630g_mmt-1755_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_630g_mmt-1755>>-
cpe:2.3:h:medtronic:minimed_630g_mmt-1755:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1711_firmware>>-
cpe:2.3:o:medtronic:minimed_640g_mmt-1711_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1711>>-
cpe:2.3:h:medtronic:minimed_640g_mmt-1711:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1712_firmware>>-
cpe:2.3:o:medtronic:minimed_640g_mmt-1712_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1712>>-
cpe:2.3:h:medtronic:minimed_640g_mmt-1712:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1751_firmware>>-
cpe:2.3:o:medtronic:minimed_640g_mmt-1751_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1751>>-
cpe:2.3:h:medtronic:minimed_640g_mmt-1751:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1752_firmware>>-
cpe:2.3:o:medtronic:minimed_640g_mmt-1752_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_640g_mmt-1752>>-
cpe:2.3:h:medtronic:minimed_640g_mmt-1752:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1740_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1740_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1740>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1740:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1741_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1741_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1741>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1741:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1742_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1742_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1742>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1742:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1760_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1760_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1760>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1760:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1761_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1761_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1761>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1761:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1762_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1762_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1762>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1762:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1780_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1780_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1780>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1780:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1781_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1781_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1781>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1781:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1782_firmware>>-
cpe:2.3:o:medtronic:minimed_670g_mmt-1782_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>minimed_670g_mmt-1782>>-
cpe:2.3:h:medtronic:minimed_670g_mmt-1782:-:*:*:*:*:*:*:*
medtronic
medtronic
>>mmt-1151_firmware>>-
cpe:2.3:o:medtronic:mmt-1151_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>mmt-1151>>-
cpe:2.3:h:medtronic:mmt-1151:-:*:*:*:*:*:*:*
medtronic
medtronic
>>mmt-1152_firmware>>-
cpe:2.3:o:medtronic:mmt-1152_firmware:-:*:*:*:*:*:*:*
medtronic
medtronic
>>mmt-1152>>-
cpe:2.3:h:medtronic:mmt-1152:-:*:*:*:*:*:*:*
Weaknesses
CWE IDTypeSource
NVD-CWE-OtherPrimarynvd@nist.gov
CWE-693Secondarysecurity@medtronic.com
CWE ID: NVD-CWE-Other
Type: Primary
Source: nvd@nist.gov
CWE ID: CWE-693
Type: Secondary
Source: security@medtronic.com
Evaluator Description

Evaluator Impact

Evaluator Solution

Vendor Statements

References
HyperlinkSourceResource
https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-600-series-communication-issue.htmlsecurity@medtronic.com
Mitigation
Vendor Advisory
https://www.cisa.gov/uscert/ics/advisories/icsma-22-263-01security@medtronic.com
Third Party Advisory
US Government Resource
Hyperlink: https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-600-series-communication-issue.html
Source: security@medtronic.com
Resource:
Mitigation
Vendor Advisory
Hyperlink: https://www.cisa.gov/uscert/ics/advisories/icsma-22-263-01
Source: security@medtronic.com
Resource:
Third Party Advisory
US Government Resource

Change History

0
Information is not available yet

Similar CVEs

2Records found

CVE-2019-13535
Matching Score-6
Assigner-Cybersecurity and Infrastructure Security Agency (CISA) Industrial Control Systems (ICS)
ShareView Details
Matching Score-6
Assigner-Cybersecurity and Infrastructure Security Agency (CISA) Industrial Control Systems (ICS)
CVSS Score-4.6||MEDIUM
EPSS-0.45% / 62.66%
||
7 Day CHG~0.00%
Published-08 Nov, 2019 | 19:45
Updated-22 May, 2025 | 19:15
Rejected-Not Available
Known To Be Used In Ransomware Campaigns?-Not Available
KEV Added-Not Available
KEV Action Due Date-Not Available
Medtronic Valleylab FT10 and LS10 Protection Mechanism Failure

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

Action-Not Available
Vendor-medtronicMedtronic
Product-valleylab_ft10_energy_platformvalleylab_ls10_energy_platformvalleylab_ls10_energy_platform_firmwarevalleylab_ft10_energy_platform_firmwareValleylab FT10 Energy Platform (VLFT10GEN)Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)
CWE ID-CWE-693
Protection Mechanism Failure
CWE ID-CWE-732
Incorrect Permission Assignment for Critical Resource
CVE-2018-10631
Matching Score-6
Assigner-Cybersecurity and Infrastructure Security Agency (CISA) Industrial Control Systems (ICS)
ShareView Details
Matching Score-6
Assigner-Cybersecurity and Infrastructure Security Agency (CISA) Industrial Control Systems (ICS)
CVSS Score-6.3||MEDIUM
EPSS-0.15% / 36.45%
||
7 Day CHG+0.02%
Published-13 Jul, 2018 | 19:00
Updated-26 Aug, 2025 | 15:15
Rejected-Not Available
Known To Be Used In Ransomware Campaigns?-Not Available
KEV Added-Not Available
KEV Action Due Date-Not Available
Medtronic N'Vision Clinician Programmer Protection Mechanism Failure

The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.

Action-Not Available
Vendor-medtronicMedtronicics_cert
Product-n\'vision_8840n\'vision_8870n\'vision_8870_firmwaren\'vision_8840_firmware8840 N’Vision Clinician Programmer8870 N’Vision removable Application Cardmedtronic_n_vision_clinician_programmer
CWE ID-CWE-693
Protection Mechanism Failure
Details not found