Logo
-

Byte Open Security

(ByteOS Network)

Log In

Sign Up

ByteOS

Security
Vulnerability Details
Registries
Custom Views
Weaknesses
Attack Patterns
Filters & Tools

Agilia Partner

Source -

CNA

CNA CVEs -

1

ADP CVEs -

0

CISA CVEs -

0

NVD CVEs -

0
Related CVEsRelated VendorsRelated AssignersReports
1Vulnerabilities found

CVE-2021-23207
Assigner-Cybersecurity and Infrastructure Security Agency (CISA) Industrial Control Systems (ICS)
ShareView Details
Assigner-Cybersecurity and Infrastructure Security Agency (CISA) Industrial Control Systems (ICS)
CVSS Score-6.5||MEDIUM
EPSS-0.09% / 26.27%
||
7 Day CHG~0.00%
Published-21 Jan, 2022 | 18:17
Updated-16 Apr, 2025 | 16:47
Rejected-Not Available
Known To Be Used In Ransomware Campaigns?-Not Available
KEV Added-Not Available
KEV Action Due Date-Not Available
Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password

An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users.

Action-Not Available
Vendor-fresenius-kabiFresenius Kabi
Product-link\+_agilia_firmwarevigilant_insightagilia_connectvigilant_mastermedlink\+_agiliavigilant_centeriumagilia_partner_maintenance_softwareVigilant Software Suite (Mastermed Dashboard)Agilia Partner
CWE ID-CWE-256
Plaintext Storage of a Password
CWE ID-CWE-522
Insufficiently Protected Credentials